How does 4-Terminal transistors work?
4
FOUR TERMINAL SENSING
Four-terminal sensing (4T sensing), 4-wire sensing, or 4-point probes method is an electrical impedance measuring technique that uses separate pairs
of current-carrying
and voltage-sensing
electrodes to make more accurate measurements than the simpler and more usual
two-terminal (2T) sensing. Four-terminal sensing is used in some ohmmeters and impedance analyzers, and in
wiring for strain gauges and resistance thermometers.
Four-point probes are also used to measure sheet resistance of thin films.
Separation of current and voltage
electrodes eliminates the lead and contact resistance from the measurement. This is an advantage
for precise measurement of low resistance values. For example, an LCR bridge instruction manual recommends the
four-terminal technique for accurate measurement of resistance below 100 ohms.[1]
Four-terminal sensing is also
known as Kelvin sensing,
after William Thomson, Lord Kelvin, who invented the Kelvin bridge in 1861 to measure very low resistances using
four-terminal sensing. Each two-wire connection can be called a Kelvin connection. A pair of
contacts that is designed to connect a force-and-sense pair to a single
terminal or lead simultaneously is called a Kelvin
contact. A clip, often acrocodile clip, that
connects a force-and-sense pair is called a Kelvin
clip.
OPERATING PRINCIPLE
When a Kelvin connection is used,
current is supplied via a pair of force connections (current leads). These
generate a voltage drop across the impedance to be measured according to Ohm's law V=RI. A pair of sense connections (voltage leads) are made
immediately adjacent to the target impedance, so that they do not include the
voltage drop in the force leads or contacts. Since almost no current flows to
the measuring instrument, the voltage drop in the sense leads is negligible.
It is usual to arrange the sense
wires as the inside pair, while the force wires are the outside pair. If the
force and sense connections are exchanged, accuracy can be affected, because
more of the lead resistance is included in the measurement. The force wires may
have to carry a large current when measuring very small resistances, and must
be of adequate gauge; the sense wires can be of a small gauge.
The technique is commonly used in
low-voltage power supplies, where it is called remote sensing, to measure the
voltage delivered to the load independent of the voltage drop in the supply
wires.
It is common to provide 4-wire
connections to current-sensing shunt
resistors of low
resistance operating at high current.
3 WIRE SENSING
A variant uses three wires, with
separate load and sense leads at one end, and a common wire on the other.
Voltage drop in the common wire is compensated for by assuming that it is the
same as in the load wire, of the same gauge and length. This technique is
widely used in resistance thermometers, also
known as resistance temperature detectors or RTDs.
Another example is in the ATX power supply standard, which includes a remote sense wire
connected to the 3.3V supply line at connector pin 13, but no sense connection
for the ground wires.
What does 4-wire testing buy you?
1. It eliminates the resistance of your interface
cabling. If fixturing resistance is a significant part of the total resistance
then using 4-wire will greatly improve accuracy.
2. It allows you to measure lower resistance
values than 2-wire testing. On Cirris hipot testers we use a higher current (up
to 1Amp) when performing 4-wire Kelvin tests. This allows us to more accurately
measure lower resistances, all the way down to 1 mΩ (0.001 ohm). Our low
voltage testers that are 4-wire capable (CR, 1100R+) can measure down to 5 mΩ ,
but can still resolve to 1 mΩ . (You lose mΩ resolution once the DUT resistance
is > 10 ohms)
3. If you make 4-wire connections on the DUT, not
just the connector that mates to the DUT, you can eliminate all sources of
fixturing resistance. This extra effort however may not be feasible.
How 4-terminal resistors work
In a 4-terminal resistor, an ultra-precise
resistor (green) is connected to 4 terminals through small, but unknown,
resistors (red).
These unknown resistors are the combination of lead resistance, screw terminal resistance, connection wire resistance, and other sources of errors.
Typical values for these unknown resistors range from 0.01 ohms to 0.2 ohms, and the values are often unstable
The values can change when you loosen or tighten a screw, for example, or if you substitute a longer test lead.
These unknown resistors are the combination of lead resistance, screw terminal resistance, connection wire resistance, and other sources of errors.
Typical values for these unknown resistors range from 0.01 ohms to 0.2 ohms, and the values are often unstable
The values can change when you loosen or tighten a screw, for example, or if you substitute a longer test lead.
To use a 4-terminal
resistor, we force a current from Terminal 1 to Terminal 2. It's current, so
the unknown resistances attached to Terminal 1 and Terminal 2 don't affect the
amount of the current. The same number of electrons per second flow through
from T1 to T2, regardless of the resistance.
A voltmeter measures the
resulting voltage drop across the ultra-precise resistor, measuring through the
unknown resistors attached to Terminal 3 and Terminal 4. The voltmeter's input
impedance is very, very high compared to the unknown resistors, so the unknown
resistors have essentially zero effect (typically less than 0.1
parts-per-million).
So the current flows
through the 0.100 ohm resistor, unaffected by the unknown resistors, and we
measure the voltage across the 0.100 ohm resistor, unaffected by the unknown
resistors.
And that's how a
4-terminal resistor works!
So what errors do we
worry about when we're using this type of resistor? We typically measure these
errors in parts-per-million, or PPM (one PPM = 0.0001%).
There are five major sources of errors: calibration uncertainty, inductance, temperature, aging, and metal-to-metal contacts.
There are five major sources of errors: calibration uncertainty, inductance, temperature, aging, and metal-to-metal contacts.
·
The United States
National Institute of Standards and Technology calibrates the PSL standard
resistor to an uncertainty of 0.5 PPM - this sets an absolute floor for
measurement uncertainty at PSL
·
The resistor is
typically made of a length of wire wrapped around a spool; in most precision
resistors, the direction of wrapping is reversed after half the turns to
minimize the inductance. Still, optimal accuracy is achieved at 0 Hertz, or DC.
Even at typical power frequencies of a few tens of hertz, the inductance can contribute
errors of a few PPM.
·
The wire is made of a
material that has minimal temperature coefficient. Still, a change of a degree
or two can contribute errors of tens to hundreds of PPM. It's not just the
ambient temperature of the oil bath, either; we have to worry about
self-heating when we pass current through the resistor. So temperature control
plays a major role
·
All resistors change
values when they age. Most studies show that there is an initial aging during
the first few years of a resistor's life (sometimes this is intentionally
accelerated with heat), then the resistor settles into a stable value. For this
reason, PSL prefers calibration resistors that are at least 25 years old
·
It's important to
consider the galvanic voltages generated by the contact between dissimilar
metals. Although the value of the unknown resistors has no effect on the
function of the calibration resistor, even a microvolt of galvanic voltage on
Terminal 3 or Terminal 4 will affect the reading by 10 PPM at 1 amp. So we're very
careful to use the correct types of test lead connectors on these terminals
(and on the terminals of the voltmeter).
Kelvin (4-wire)
resistance measurement
Any voltage dropped across the main
current-carrying wires will not be measured by the voltmeter, and so do not
factor into the resistance calculation at all. Measurement accuracy may be
improved even further if the voltmeter's current is kept to a minimum, either
by using a high-quality (low full-scale current) movement and/or a
potentiometric (null-balance) system.
This method of measurement
which avoids errors caused by wire resistance is called the Kelvin, or 4-wire method. Special connecting clips called Kelvin clips are made to facilitate this kind of
connection across a subject resistance:
In regular,
"alligator" style clips, both halves of the jaw are electrically
common to each other, usually joined at the hinge point. In Kelvin clips, the
jaw halves are insulated from each other at the hinge point, only contacting at
the tips where they clasp the wire or terminal of the subject being measured.
Thus, current through the "C" ("current") jaw halves does
not go through the "P" ("potential," or voltage) jaw halves, and will
not create any error-inducing voltage drop along their length:
The same
principle of using different contact points for current conduction and voltage
measurement is used in precision shunt resistors for measuring large amounts of
current. As discussed previously, shunt resistors function as current
measurement devices by dropping a precise amount of voltage for every amp of
current through them, the voltage drop being measured by a voltmeter. In this
sense, a precision shunt resistor "converts" a current value into a
proportional voltage value. Thus, current may be accurately measured by
measuring voltage dropped across the shunt:
Current measurement using a shunt
resistor and voltmeter is particularly well-suited for applications involving
particularly large magnitudes of current. In such applications, the shunt
resistor's resistance will likely be in the order of milliohms or microohms, so
that only a modest amount of voltage will be dropped at full current.
Resistance this low is comparable to wire connection resistance, which means
voltage measured across such a shunt must be done so in such a way as to avoid
detecting voltage dropped across the current-carrying wire connections, lest
huge measurement errors be induced. In order that the voltmeter measure only
the voltage dropped by the shunt resistance itself, without any stray voltages
originating from wire or connection resistance, shunts are usually equipped
with four connection terminals:
In metrological (metrology = "the
science of measurement") applications, where accuracy is of paramount
importance, highly precise "standard" resistors are also equipped
with four terminals: two for carrying the measured current, and two for
conveying the resistor's voltage drop to the voltmeter. This way, the voltmeter
only measures voltage dropped across the precision resistance itself, without
any stray voltages dropped across current-carrying wires or wire-to-terminal
connection resistances.
The following photograph shows a
precision standard resistor of 1 Ω value immersed in a temperature-controlled
oil bath with a few other standard resistors. Note the two large, outer
terminals for current, and the two small connection terminals for voltage:
The Kelvin measurement can be a practical
tool for finding poor connections or unexpected resistance in an electrical
circuit. Connect a DC power supply to the circuit and adjust the power supply
so that it supplies a constant current to the circuit as shown in the diagram
above (within the circuit's capabilities, of course). With a digital multimeter
set to measure DC voltage, measure the voltage drop across various points in
the circuit. If you know the wire size, you can estimate the voltage drop you
should see and compare this to the voltage drop you measure. This can be a
quick and effective method of finding poor connections in wiring exposed to the
elements, such as in the lighting circuits of a trailer. It can also work well
for unpowered AC conductors (make sure the AC power cannot be turned on). For
example, you can measure the voltage drop across a light switch and determine
if the wiring connections to the switch or the switch's contacts are suspect.
To be most effective using this technique, you should also measure the same
type of circuits after they are newly made so you have a feel for the
"correct" values. If you use this technique on new circuits and put
the results in a log book, you have valuable information for troubleshooting in
the future.
Comments
Post a Comment